T-Scan Digital Occlusal Analysis in Practice

T-Scan adds timing and relative force to the occlusal check. What the sensor measures, how to use it when adjusting restorations, and what the evidence says about accuracy.

November 17, 20269 min read1,609 words
White gypsum model of both dental arches in occlusion, side view, on a dark grey lab bench in cool light
Contents
  1. What articulating paper cannot show
  2. The T-Scan system in brief
  3. What the evidence says
  4. Where T-Scan earns its place
  5. A chairside sequence
  6. Reading the numbers
  7. Limits to keep in mind
  8. Summary
  9. Frequently asked questions

T-Scan is a digital occlusal analysis system that records the order in which teeth meet and how the bite force is shared between them. Articulating paper cannot tell you either. This article explains what the T-Scan Novus sensor measures, how to build it into occlusal adjustment of provisionals and definitive restorations, which reference values exist, and where the method reaches its limits.

What articulating paper cannot show

A paper or film mark tells you that two surfaces touched. It does not tell you how hard they touched, or which contact came first. Mark size is a poor guide to force: Carey and colleagues (2007) found no direct relationship between mark area and applied load, and in a clinical study of 30 women the largest mark belonged to the tooth with the highest force only 38.3 % of the time (Qadeer et al. 2012). How to choose and use paper, film, silk and shimstock is covered in articulating paper and shimstock.

Digital occlusal analysis adds two dimensions:

  • Time: the first contact, the time from first contact to full intercuspation, and how quickly posterior teeth separate in excursions.
  • Relative force: the percentage of total bite force carried by each tooth, each side and each segment.

Dr Hajtó teaches these two dimensions as the real value of T-Scan, whatever occlusal concept you are aiming for. They make premature contacts and overloaded teeth visible before you start grinding.

The T-Scan system in brief

Tekscan currently offers two platforms: T-Scan Novus, the full system, and T-Scan Novus Core, which the manufacturer describes as the entry-level system for general practice. Both use the same sensors and the same handpiece, which connects to a Windows computer by USB.

FeatureT-Scan Novus (Tekscan)OccluSense (Bausch)
Sensor thickness100 µm60 µm
Sensing points (sensels)2,2001,088
Force levels256256
Sensor sizes22
Time resolution0.003-second incrementstime curve
Marks the teethnoyes, red colour coating

These are the figures Dr Hajtó compares in his occlusion courses. For T-Scan, Kerstein and Radke (2022) confirm a 100 µm sensor with 2,200 sensels and 256 force levels; Bausch states 60 µm and 256 pressure levels for OccluSense. Sensitivity is adjustable on both.

According to Tekscan, a sensor is reusable on one patient, withstands roughly 15 to 25 closures and is cold-sterilised between visits. Software features include implant loading alerts, overlay with STL files from an intraoral scan, a centre-of-force trajectory, a timing table and video export.

What the evidence says

The literature is large and not uniform. Grouped by question:

Is it repeatable? In 42 subjects, Koos and colleagues (2010) found a measurement error of 1 %, an accuracy of 2 % and a reliability of 2.8 %, with no significant effect of changing the sensor or repeating the measurement. Stern and Kordaß (2010) reported 80.5 % agreement with the Greifswald Digital Analyzing System in 38 subjects. In vitro, T-Scan measurements of left-right force balance were far more repeatable than OccluSense: repeatability and reproducibility accounted for 0.8 % of measurement variance with T-Scan and 12 % with OccluSense (Jauregi et al. 2024).

Does it measure real force? No, it measures relative force. When Cerna and colleagues loaded 18 sensors with known forces, readings were highly repeatable but not valid as absolute total force.

Does it find every contact? Not quite. Reich and colleagues (2025) compared articulating silk, a wax occlusogram and T-Scan Novus in 26 adults. Per arch, silk recorded 15.3 contacts on average, wax 14.2 and T-Scan 9.4. Agreement was best for incisors and canines and weakest for premolars and molars. T-Scan, however, showed clearly higher intra- and inter-observer reliability than the wax occlusogram, and the authors suggest it when force distribution is the main question.

Overall: a systematic review of 86 studies concluded that T-Scan measures relative contact forces and timing objectively and repeatably (Chowdhary and Sonnahalli 2024). A 2026 Dutch review stresses conflicting results on contact location and regards articulating paper as still essential (Mir et al. 2026). For OccluSense, agreement with shimstock in typodonts was only fair (κ 0.22) against 0.94 for 40 µm paper (Schütze et al. 2025).

The practical conclusion matches what Dr Hajtó has taught for years: use the sensor and the paper together. Because the sensor leaves no mark on the teeth, Koos and colleagues reached the same conclusion.

Hands-on dental courses

Courses for dentists in Munich, several of them taught in English.

See all courses

Where T-Scan earns its place

In Dr Hajtó's workflow the sensor is used whenever many contacts have to be right at the same time:

  • adjusting full-arch long-term provisionals in a full mouth rehabilitation,
  • adjusting definitive all-ceramic restorations after cementation,
  • checking and fine-tuning bonded occlusal build-ups when the vertical dimension of occlusion is raised; the build-ups may be made slightly higher than the final vertical so that there is material left to adjust.

For the overall sequence of such cases, see full mouth rehabilitation of the worn dentition.

A chairside sequence

  1. Choose the sensor size and insert it in the handpiece. The support centres the sensor between the upper central incisors (11 and 21).
  2. Set up the patient file and adapt the arch in the software, which the manufacturer describes as customisable, so that forces are assigned to the right teeth.
  3. Adjust sensitivity to the patient; it is variable.
  4. Record a baseline: several closures into maximum intercuspation, then protrusion and lateral excursions.
  5. Analyse the first contact, the time to full closure, the force share per tooth and side, and disclusion in excursions. The recording can be played back frame by frame.
  6. Locate the offending contact on the tooth with articulating paper or film.
  7. Adjust and re-record until timing and force distribution are acceptable. Repeat recordings in the same visit are unlimited.
  8. Save baseline and final recordings for the patient record.

Reading the numbers

Reference values for healthy dentitions now exist. Rubió-Ferrer and colleagues (2024) recorded 178 adults:

ParameterHealthy range
Force share, right side (5th to 95th percentile)34 to 67 %
Force share, posterior teeth (5th to 95th percentile)55 to 94 %
Occlusion time, multi-bite (90th percentile)0.17 s
Occlusion time, single bite (90th percentile)0.27 s

Reliability of these parameters was only moderate, with intraclass correlations of 0.58 to 0.70. A single recording is therefore not a diagnosis. Record several closures and interpret the values with the clinical findings. Dr Hajtó compares a T-Scan recording to a radiograph: both need interpretation.

Timing data are also used in research on non-carious cervical lesions. Haralur and colleagues (2019) found longer occlusion times (0.727 s against 0.516 s) and longer disclusion times in all excursions in 50 patients with such lesions than in 50 controls. That is an association, not proof of cause; see abfraction lesions and NCCLs.

Limits to keep in mind

  • Relative, not absolute: percentages say nothing about the patient's actual bite force.
  • Sensor in the way: a 100 µm sensor sits between the teeth, so the recorded closure is not identical to closure without it.
  • Fewer contacts: in the clinical comparison T-Scan registered fewer contacts than silk and wax, and agreement between methods was weakest for premolars and molars.
  • No marks: without paper you do not know which cusp slope to adjust.
  • No concept: the sensor does not decide which contacts and guidance you want. That is set during planning, for example with canine guidance.

Patients who keep reporting a wrong bite despite unremarkable findings need a different approach from repeated adjustment; see occlusal dysesthesia. German-speaking readers will find Dr Hajtó's earlier practice report in Präziser Biss mit dem T-Scan.

Summary

T-Scan shows the sequence and relative distribution of occlusal forces, is highly repeatable, and complements articulating paper rather than replacing it. It is most useful when adjusting extensive provisional and definitive restorations. Occlusion concepts, articulation and bite raising are taught in the English-language VDO and occlusion course in Munich.

Frequently asked questions

What does T-Scan measure?

T-Scan measures the timing and relative force of occlusal contacts. A thin pressure sensor records each closure, and the software shows which tooth contacts first, how long it takes to reach full intercuspation and what percentage of the total force each tooth and side carries. It does not measure absolute bite force, and the sensor leaves no marks on the teeth.

Is T-Scan more accurate than articulating paper?

It answers a different question. Paper shows where teeth touch; T-Scan shows when and how hard relative to other teeth. Studies report good repeatability, for example a 1 % measurement error in 42 subjects. In a clinical comparison, however, T-Scan registered fewer contacts per arch than articulating silk, so it does not replace a marking medium.

How many times can a T-Scan sensor be used?

According to the manufacturer, a sensor is reusable on a single patient and withstands roughly 15 to 25 closures. It is cold-sterilised between visits. Keeping the patient's sensor allows follow-up recordings with the same sensor, which makes before-and-after comparisons more consistent.

What is the difference between T-Scan Novus and Novus Core?

Both are current Tekscan platforms and use the same sensors and handpiece. Novus Core is described by the manufacturer as the entry-level system for general dental practice, while Novus offers the full feature set, including tools such as implant loading alerts and overlay with intraoral scan data. The choice depends on which analyses you need.

What is OccluSense and how does it compare?

OccluSense by Bausch is a 60 µm sensor with 256 pressure levels and a red coating that also marks contacts on the teeth. In an in vitro study its left-right balance measurements were less repeatable than T-Scan's, and in typodonts its agreement with shimstock was only fair (κ 0.22), particularly poor in deep bite cases.